Dual Decontamination Mixing Valve System with Temperature Override Protection
TM-855
Add to favorites Submit RFQ Now
PDF Submittal Data Sheet
PDF Operation and Maintenance
PDF Maintenance Guide
PRIMARY THERMOSTATIC WATER MIXING VALVE  
• DURA-trol® solid bimetal thermostat directly linked to valve porting to control the intake of hot and cold water and compensate for supply temperature and pressure fluctuations.  
• Primary valve can be set to the correct temperature for the application.  
• Locking type temperature regulator to prevent accidental movement  
• Adjustable high temperature limit stop set for 120ºF (49ºC)  
• Full 1 1/4" top inlets and 1 1/4" top outlet  
• Dial thermometer (range 0 to 140°F, -10 to 60°C)  
• Rough bronze finish  
• Union angle checkstops on inlets  
 
REDUNDANT THERMOSTATIC CONTROL VALVE  
• Stainless steel bellows adjustable thermostat to allow cold water to enter the outlet side of the Primary mixing valve.  
• Remains fully closed until outlet temperature reaches set point.  
• Will not allow set point temperature to be exceeded.  
• Maximum supply temperature 180ºF (82ºC)  
• Maximum supply pressure 125 PSI (8.6 Bar)

< Go Back

Specification
Inlets 1"
Outlet 1 1/4"
Minimum Flow(GPM) 3
Maximum Flow(GPM) 64
Weight (pounds) 40